Table 1 Inter equipment comparison between FACSCalibur and EPICS XL-MCL using the stabilized blood samples

From: Establishment of reference CD4+ T cell values for adult Indian population

  Immuno Trol Lot# 7587024 (N = 30) UKNEQAS proficiency samples(N = 22)
  r value %CV r value %CV
Absolute CD4 count(cells/μL) 0.917 5.17 0.975 7.19
CD4 percentage 0.891 7.04 0.993 4.6
  1. The inter equipment comparison between FACSCalibur and EPICS XL-MCL was conducted using the stabilized blood samples. The r values (range: 0.89 to 0.99) were acceptable. The % CV (range: 4.6 to 7.1) were > 10% and hence acceptable.