Skip to main content

Table 1 IQC details of CD4 testing laboratories

From: Duplicate analysis method: a cheaper alternative to commercial IQC materials in limited resource settings for monitoring CD4 testing

Laboratories

Instrument used

Mean (range of % variation for duplicate analysis)

Mean (range % CV for commercially available controls)

Laboratory 1

FACSCount

2.43, 0.6–3.3

3.9, 2.7–4.9 (IMMUNO-TROL)

Laboratory 2

FACSCount

7.2, 5.5–12.0

5.4, 3.4–7.0 (multi-check)

Laboratory 3

FACSCalibur

5.7, 2.5–7.2

12.4, 12.0–12.8 (multi-check)

Laboratory 4

FACSCalibur

5.1, 4.2–5.8

–

Laboratory 5

FACSCalibur

4.4, 2.6–6.4

2.2, 1.6–2.6 (multi-check)

Laboratory 6

FACSCalibur

5.9, 4.2–7.0

3.4, 3.0–3.8 (multi-check)

Laboratory 7

FACSCount

2.6, 2.1–3.5

3.5 (multi-check)

Laboratory 8

Cyflow

2.9, 2.0–3.9

–

Laboratory 9

Cyflow

0.5, 0.3–1.2

–

Laboratory 10

Cyflow

3.5, 1.6–5.6

–

Laboratory 11

FACSCalibur

4.8, 3.5–5.3

–